Specifications CL3436
This transmitter powered with a dc voltage between 9 and 36 volt provides an isolated 4-20 mA "current loop", a RS485 and Modbus (function 03, 06, 16) outputs.
The instrument works with potentistatic or polarogaphic cell and RTD Pt100 temperature sensors.
It provides manual and automatic temperature compensation. If an RTD is connected, the instruments provides the temperature reading.
The temperature coefficient is selectable by the operator.
Its modularity, small dimensions and built-in technology make it suitable for measurements of oxidizers in various process applications.
Features CL3436
Scales: 0 ÷ 2,000 / 0 ÷ 20.00 / 0 ÷ 200.0 ppm - mg/l
Temperature scales: -10.0 ÷ 110.0 °C / 14.0 ÷ 230.0 °F
Digital output: RS485 isolated
Analog output: 4-20 mA isolated, "current loop" two wires, scalable
Power supply: 9 / 36 Vdc
Dimensions: 71x95x58 mm. Rail Din (4 modules)
EMC/RFI conformity: EN 61326
Registered design: 002564666-001